DocumentCode
2119623
Title
Micro preconcentrator for handheld diagnostics of cancer biomarkers in breath
Author
Alfeeli, Bassam ; Agah, Masoud
Author_Institution
Dept. of Electr. & Comput. Eng., Kuwait Inst. for Sci. Res., Kuwait City, Kuwait
fYear
2010
fDate
1-4 Nov. 2010
Firstpage
2490
Lastpage
2493
Abstract
This paper introduces, for the first time, Maltese-cross pillar design implemented in a 7 mm × 7 mm × 0.35 mm cavity for high-performance microfabricated preconcentrators (μPCs). This high-aspect ratio design provides more surface area over the previously reported basic geometric shapes. Results showed that Maltese-cross provided 27% increases in performance. Moreover, reducing the pillar spacing from 100 μm to 10 μm and arranging the pillars in staggered fashion have resulted in 145% and 85% increase in performance, respectively. Therefore, it was concluded that both surface area and flow pattern contribute significantly to the device performance. These μPCs achieved high adsorption capacity making them suitable for handheld breath analyzers to concentrate and identify trace-level (low ppm) volatile compounds (cancer biomarkers) found in human breath.
Keywords
bioMEMS; cancer; chromatography; microfluidics; patient diagnosis; Maltese-cross pillar design; adsorption capacity; cancer biomarkers; distance 10 mum; distance 100 mum; handheld breath analyzers; handheld diagnostics; micropreconcentrator; size 0.35 mm; size 7 mm; trace-level volatile compounds;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2010 IEEE
Conference_Location
Kona, HI
ISSN
1930-0395
Print_ISBN
978-1-4244-8170-5
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2010.5690098
Filename
5690098
Link To Document