    Different defibrillation waveforms exhibit different efficacies. It has been proposed that a resistor-capacitor (RC) model of the myocardial membrane can explain these differences. Both the maximum capacitor voltage (charge banking) and the remaining voltage after the pulse (charge burping) have been proposed as efficacy predictors. In this paper, we test these hypotheses against 140 defibrillation waveforms. Data is reported from 159 guinea pigs. For each tested waveform, we calculated to within a fixed constant the minimum and maximum voltage on the capacitor, and the voltage remaining after the shock for time constants (τ) between 0.1 ms and 4.0 ms. Each parameter was tested to determine if it would be predicted to have a 10% impact on defibrillation efficacy. No parameter was a significant predictor of defibrillation efficacy for τ>2.0 ms. The maximum voltage was significant for short time constants (τ<=2.0 ms, p<0.005). For the very short time constants (<=0.5 ms) the voltage remaining was significant (p<0.05), and, for the shortest time constants (t=0.1 ms), the minimum voltage was a significant predictor (p<0.05)
  • Keywords
    RC circuits; bioelectric potentials; biomembranes; capacitors; cardiology; defibrillators; electric shocks; neuromuscular stimulation; waveform analysis; 0.1 to 4.0 ms; charge banking; charge burping; defibrillation mechanisms; defibrillation waveform efficacy prediction; electric shock; guinea pigs; maximum capacitor voltage; minimum voltage; remaining voltage; resistor-capacitor myocardial membrane model; time constants; Animals; Biomembranes; Capacitors; Cells (biology); Defibrillation; Electric shock; Electrodes; Predictive models; Testing; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 2001
  • Conference_Location
    Rotterdam
  • ISSN
    0276-6547
  • Print_ISBN
    0-7803-7266-2
  • Type

    conf

  • DOI
    10.1109/CIC.2001.977637
  • Filename
    977637