• DocumentCode
    2126838
  • Title

    International comparison for RF power in the frequency range up to 18 GHz

  • Author

    de Vreede, J.P.M. ; Korfage-Janssen, W.

  • Author_Institution
    NMi Van Swinden Lab., Delft, Netherlands
  • fYear
    2000
  • fDate
    14-19 May 2000
  • Firstpage
    58
  • Lastpage
    59
  • Abstract
    A Euromet international laboratory comparison (Project 393) has been carried out between 15 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The results show good agreement in measuring the calibration coefficient of the DUTs and in measuring their reflection coefficients.
  • Keywords
    Calibration; Measurement uncertainty; Microwave measurement; Power measurement; Thermistors; Transfer standards; UHF measurement; 10 MHz to 18 GHz; Euromet project; RF power measurement; calibration coefficient; international laboratory comparison; national standards laboratories; reflection coefficients; thermistor mounts; travelling scheme; uncertainties; Calibration; Connectors; Hafnium; Laboratories; Metrology; Particle measurements; Radio frequency; Reflection; Telecommunications; Thermistors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2000 Conference on
  • Conference_Location
    Sydney, NSW, Australia
  • Print_ISBN
    0-7803-5744-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2000.850875
  • Filename
    850875