Abstract :
The traditional scaling method used at IEN for the calibration of the 10 k/spl Omega/ standard is compared with more straightforward techniques, one based on the linearity of a DVM and the other on a commercial current comparator bridge. The results of the three methods agree to better than 1/spl middot/10/sup -7/.
Keywords :
Calibration; Electric resistance measurement; Measurement standards; Measurement uncertainty; Quantum Hall effect; 10 kohm; DVM linearity; QHE; calibration; current comparator bridge; resistance standard; scaling method; traceability; uncertainty budgets; Bridge circuits; Calibration; Current supplies; Laboratories; Linearity; Measurement standards; Performance evaluation; Resistors; Thermal resistance; Voltage;