  • Abstract
    The NIST dual-balance guarded active-arm bridge, using a third dc source and a second detector to balance the guard network, is described. Improvements to the NIST guarded active-arm bridge and the design and construction of improved high resistance standards have facilitated the extension of NIST calibration services to the 10 T/spl Omega/ and 100 T/spl Omega/ decades of resistance.
  • Keywords
    bridge instruments; calibration; electric resistance measurement; measurement standards; resistors; 1 T/spl Omega/ to 100 T/spl Omega/; 1 to 100 Tohm; NIST calibration services; construction; dc source; design; detector; dual-balance high-resistance bridge; guarded active-arm bridge; Bridge circuits; Calibration; Detectors; Laboratories; Leakage current; Measurement standards; Monitoring; Resistors; US Department of Commerce; Voltage;