DocumentCode :
2129696
Title :
Calibration of a standard resistance using a standard platinum resistance thermometer
Author :
Valencia-Rodriguez, J. ; Figueroa-Estrada, J.M. ; Escamilla-Esquivel, A. ; Martinez-Martinez, S. ; Martinez-Guerrero, B.
Author_Institution :
Centro de Investigacion de Ciencia Avanzada y Technol. Aplicada, Mexico
fYear :
2000
fDate :
14-19 May 2000
Firstpage :
307
Lastpage :
308
Abstract :
A calibrated standard resistance (R/sub s/)is indispensable in order to perform reliable temperature measurements with a Standard Platinum Resistance Thermometer (SPRT). Since in many instances the calibration of such R/sub s/ is difficult to achieve, a novel method for such calibration using two SPRTs, a bridge and a bath is proposed.
Keywords :
bridge instruments; calibration; measurement standards; platinum; resistance thermometers; temperature measurement; Pt; Pt resistance thermometer; bath; bridge; calibration; standard resistance; temperature measurements; Bridges; Calibration; Electrical resistance measurement; Laboratories; Measurement standards; Performance evaluation; Platinum; Resistors; Temperature; Thermal resistance;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2000 Conference on
Conference_Location :
Sydney, NSW, Australia
Print_ISBN :
0-7803-5744-2
Type :
conf
DOI :
10.1109/CPEM.2000.850999
Filename :
850999
Link To Document :
بازگشت