• DocumentCode
    2129922
  • Title

    On the choice of comparison reference values for the purpose of pair-wise comparison of laboratories

  • Author

    White, D.R.

  • Author_Institution
    Measurement Standards Lab., Lower Hutt, New Zealand
  • fYear
    2000
  • fDate
    14-19 May 2000
  • Firstpage
    325
  • Lastpage
    326
  • Abstract
    A reference value is sought that provides unbiased estimates of the differences between any pair of laboratories. It is found that in general no single-valued reference value exists that yields an unbiased estimate of the difference for all possible pairs of laboratories. In all cases the best estimate of laboratory difference is evaluated strictly in terms of differences in results. It is concluded that the concept of a comparison reference value is both unnecessary and unsuited to the purpose of pair-wise comparisons of laboratories.
  • Keywords
    calibration; laboratories; measurement standards; measurement uncertainty; statistical analysis; artefacts; comparison reference values; laboratory differences; mathematical model; pair-wise comparison; statistical properties; unbiased estimates; variance; weighted sums; Joining processes; Laboratories; Mathematical model; Measurement standards; Particle measurements; Protocols; Solids; Yield estimation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2000 Conference on
  • Conference_Location
    Sydney, NSW, Australia
  • Print_ISBN
    0-7803-5744-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2000.851008
  • Filename
    851008