Title : 
Improved modelling of the coaxial Kelvin balance
         
        
            Author : 
Wood, B.M. ; Cote, M.
         
        
            Author_Institution : 
Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, Ont., Canada
         
        
        
        
        
        
            Abstract : 
We present the results of a model of the Kelvin balance commonly used in four-terminal coaxial bridges. A Kelvin balance is used to combine the low-potential terminals of two impedances into a single potential. The model considers the effect of interconnecting cables, equalisers, and shunt impedances within the measured impedances.
         
        
            Keywords : 
bridge instruments; electric impedance measurement; equalizers; multiterminal networks; coaxial Kelvin balance; equalisers; four-terminal coaxial bridges; interconnecting cables; low-potential terminals; shunt impedances; Bridge circuits; Cables; Capacitance; Coaxial components; Conductors; Detectors; Electrical resistance measurement; Equalizers; Impedance measurement; Kelvin;
         
        
        
        
            Conference_Titel : 
Precision Electromagnetic Measurements Digest, 2000 Conference on
         
        
            Conference_Location : 
Sydney, NSW, Australia
         
        
            Print_ISBN : 
0-7803-5744-2
         
        
        
            DOI : 
10.1109/CPEM.2000.851058