DocumentCode
2133073
Title
Neuromuscular electrical stimulation using different waveforms: Properties comparison by applying single pulses
Author
Meza-Cuevas, Mario A. ; Schroeder, Damien ; Krautschneider, Wolfgang H.
Author_Institution
Inst. of Nanoelectron., Hamburg Univ. of Technol., Hamburg, Germany
fYear
2012
fDate
16-18 Oct. 2012
Firstpage
840
Lastpage
845
Abstract
Electrical Neurostimulation is useful in several medical therapies, for the treatment of neuromuscular deficits and for restoring physiological and sensory functions. Fully implantable devices are preferable in order to reach high selectivity and to give comfort to the patient. However, smaller devices bring more challenges regarding fabrication techniques and technology compliances, and power consumption is an important issue. In this study we analyze the use of different waveforms for performing neuromuscular stimulation, and we show how different waveforms require less energy to produce the same muscle reaction.
Keywords
mechanoception; medical disorders; neuromuscular stimulation; patient rehabilitation; physiology; prosthetics; waveform analysis; fabrication technique; implantable device; medical therapy; muscle reaction; neuromuscular deficit treatment; neuromuscular electrical stimulation; neuromuscular stimulation; physiological function restoration; power consumption; sensory function restoration; single pulse; technology compliances; waveform; electical stimulation; functional neuromuscular stimulation; neurostimulation; stimulation waveform;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering and Informatics (BMEI), 2012 5th International Conference on
Conference_Location
Chongqing
Print_ISBN
978-1-4673-1183-0
Type
conf
DOI
10.1109/BMEI.2012.6512988
Filename
6512988
Link To Document