DocumentCode :
2133461
Title :
Evaluation of major frequency shifts in an optically pumped cesium beam frequency standard
Author :
Young Ho Park ; Sung Hoon Yang ; Ho Seong Lee ; Sang Eon Park ; Hyuck Cho
Author_Institution :
Time & Frequency Lab., Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
fYear :
2000
fDate :
14-19 May 2000
Firstpage :
629
Lastpage :
630
Abstract :
The accuracy evaluation of the optically pumped cesium beam frequency standard with a 37-cm long Ramsey cavity, KRISS-1, is being performed. Some major frequency shifts are evaluated; e.g., second-order Zeeman and Doppler shifts, cavity pulling, Rabi pulling, etc. The overall uncertainty is expected to be at the low 10/sup -13/ level in the KRISS-1.
Keywords :
Doppler shift; Zeeman effect; caesium; frequency standards; laser frequency stability; measurement by laser beam; measurement uncertainty; optical pumping; 37 cm; Cs; Cs beam frequency standard; Doppler shifts; KRISS-1; Rabi pulling; Ramsey cavity; accuracy; cavity pulling; frequency shifts; measurement uncertainty; optically pumped frequency standard; second-order Zeeman shifts; Atom optics; Atomic beams; Frequency synthesizers; Laser excitation; Magnetic fields; Optical beams; Optical pumping; Servomechanisms; Temperature; Uncertainty;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2000 Conference on
Conference_Location :
Sydney, NSW, Australia
Print_ISBN :
0-7803-5744-2
Type :
conf
DOI :
10.1109/CPEM.2000.851168
Filename :
851168
Link To Document :
بازگشت