Title
EMFi material as wearable heart rate sensor for night time recordings
Author
Vehkaoja, Antti ; Salpavaara, Timo ; Verho, Jarmo ; Lekkala, Jukka
Author_Institution
Dept. of Autom. Sci. & Eng., Tampere Univ. of Technol., Tampere, Finland
fYear
2010
fDate
1-4 Nov. 2010
Firstpage
145
Lastpage
149
Abstract
We propose a method for night time monitoring of heart rate. Our method uses one or more pressure sensitive elements integrated into a wristwatch. The sensing elements are made of film type electret material called EMFi (Electro Mechanical Film). We designed the signal processing algorithms keeping in mind on-line processing and later integration into a micro controller for wearable applications. We have evaluated our method by comparing beat-to-beat heart rate values obtained with EMFi sensors with accurate reference values calculated from reference ECG signal. We suggest based on the 209 s average continuous recognition time and 0.51 bpm mean error that our method provides adequate accuracy for night time heart rate and heart rate variability analysis, which enables the method to be used in various applications.
Keywords
biomedical equipment; biomedical materials; biomedical measurement; electrocardiography; medical signal processing; microcontrollers; patient monitoring; sensors; ECG signal; EMFi material; EMFi sensors; electret material; electro mechanical film; heart rate variability analysis; microcontroller; night time monitoring; night time recordings; on-line processing; pressure sensitive elements; wearable applications; wearable heart rate sensor; wristwatch;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2010 IEEE
Conference_Location
Kona, HI
ISSN
1930-0395
Print_ISBN
978-1-4244-8170-5
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2010.5690674
Filename
5690674