DocumentCode :
2137179
Title :
Low cost thermal measurement of pulse-transit-time
Author :
James, M. ; Gregori, S. ; Fayek, D.
Author_Institution :
Sch. of Eng., Univ. of Guelph, Guelph, ON
fYear :
2008
fDate :
4-7 May 2008
Abstract :
This work lays out our investigation into the non-invasive, automatic measurement of blood pressure. We have investigated the use of a thermal camera to detect changes in surface temperature due to the pulse in the radial and ulner arteries and then moved on to detecting the same pulse using a low-cost thermistor. Our preliminary results presented here indicate that the pulse is indeed detectable using a simple thermistor apparatus and that it is possible to develop a calibrated system to measure blood pressure using the pulse transit time method.
Keywords :
biothermics; blood pressure measurement; blood vessels; infrared imaging; noninvasive automatic blood pressure measurement; pulse transit time thermal measurement; radial artery; surface temperature changes; thermal camera; thermistor; ulner artery; Biomedical monitoring; Blood flow; Blood pressure; Catheters; Costs; Heat transfer; Pressure measurement; Pulse measurements; Thermistors; Time measurement; PTT; blood pressure; correlation; pulse transit time; thermistor;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electrical and Computer Engineering, 2008. CCECE 2008. Canadian Conference on
Conference_Location :
Niagara Falls, ON
ISSN :
0840-7789
Print_ISBN :
978-1-4244-1642-4
Electronic_ISBN :
0840-7789
Type :
conf
DOI :
10.1109/CCECE.2008.4564846
Filename :
4564846
Link To Document :