Title : 
Needle placement under X-ray fluoroscopy using perspective invariants
         
        
            Author : 
Bascle, Benedicte ; Navab, Nassir ; Loser, Michael ; Geiger, Bernhard ; Taylor, Russell
         
        
            Author_Institution : 
Imaging & Visualization Dept., Siemens Corp. Res. Inc., Princeton, NJ, USA
         
        
        
        
        
        
            Abstract : 
Presents an approach for image-based guidance of a surgical tool towards multiple targets from fixed or variable entry points. The method is based on visual servoing. It requires no prior calibration or registration. By taking advantage of projective invariants, precise needle alignment to a target can be achieved in a fixed number (12) of iterations. Alignment to n targets can be performed in 6*(n+1) iterations. Elements of error analysis and a discussion of the “optimal” placement of the planes used in the method are given. The authors also show how the approach can be used to estimate the entry point and orientation to reach an anatomic target while passing through a given anatomic landmark
         
        
            Keywords : 
biomedical equipment; diagnostic radiography; iterative methods; medical image processing; surgery; X-ray fluoroscopy; error analysis elements; fixed iterations number; image-based guidance; needle placement; optimal planes placement; percutaneous procedures; perspective invariants; projective invariants; surgical tool; targets alignment; visual servoing; Biomedical engineering; Biomedical imaging; Needles; Optical imaging; Robots; Surgery; Surges; Visual servoing; Visualization; X-ray imaging;
         
        
        
        
            Conference_Titel : 
Mathematical Methods in Biomedical Image Analysis, 2000. Proceedings. IEEE Workshop on
         
        
            Conference_Location : 
Hilton Head Island, SC
         
        
            Print_ISBN : 
0-7695-0737-9
         
        
        
            DOI : 
10.1109/MMBIA.2000.852359