Title :
Design of phased array coils for increasing the signal-to-noise ratio of magnetic resonance imaging
Author :
Liang, Dandan ; Hui, Hon Tat ; Yeo, Tat Soon
Author_Institution :
Dept. of Electr. & Comput. Eng., Nat. Univ. of Singapore, Singapore, Singapore
Abstract :
This paper introduces a method for increasing the signal-to-noise ratio (SNR) for magnetic resonance imaging (MRI) by phased array coils. The SNR increase is first theoretically explained and then demonstrated in a simulated MRI system. It is shown by simulation results that the SNR can be increased with the number of coils in the array by properly combining the output of each coil element, and the combiner coefficients are determined by decoupling matrix.
Keywords :
biomedical MRI; image resolution; matrix algebra; medical image processing; SNR; coil element; combiner coefficients; decoupling matrix; image resolution; magnetic resonance imaging; phased array coils; signal-to-noise ratio; simulated MRI system; Arrays; Circuit noise; Coils; Magnetic resonance imaging; Mutual coupling; Signal to noise ratio;
Conference_Titel :
Antennas and Propagation Society International Symposium (APSURSI), 2012 IEEE
Conference_Location :
Chicago, IL
Print_ISBN :
978-1-4673-0461-0
DOI :
10.1109/APS.2012.6349160