• DocumentCode
    2176317
  • Title

    Abnormal Acoustic Noise Reduction in X-Ray Tube Single-Phase Motor

  • Author

    Luo, Weitao ; Wang, Xu

  • Author_Institution
    Sino-Dutch Biomed. & Inf. Eng. Sch., Northeastern Univ. Shenyang, Shenyang, China
  • fYear
    2009
  • fDate
    17-19 Oct. 2009
  • Firstpage
    1
  • Lastpage
    7
  • Abstract
    This paper presents series of novel technology for reducing abnormal acoustic noises in the medical-use X-ray tube single-phase motor. Single-phase H-bridge power converter combined with phase shift control, soft boost and single-polar modulate technology was used to implement the purpose. The experimental results show that the maximum level of acoustic tube running noises is 43.5dB if the proposed scheme is used when a 3kW single-phase motor was operated from 0 to 10800 rpm. On the contrary, the level of running abnormal acoustic noises reaches 51.2dB if the conventional method is employed. This result shows that the unsuitable boost mode and working frequency of the motor are the major noise source and the method proposed in this paper is effective for the noise reduction in this kind of motor.
  • Keywords
    AC motors; X-ray tubes; noise abatement; power convertors; X-ray tube single-phase motor; abnormal acoustic noise reduction; phase shift control; single-phase H-bridge power converter; single-polar modulate technology; Acoustic noise; Anodes; Biomedical imaging; Frequency; Insulated gate bipolar transistors; Noise reduction; Noise shaping; Resonance; Stators; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics, 2009. BMEI '09. 2nd International Conference on
  • Conference_Location
    Tianjin
  • Print_ISBN
    978-1-4244-4132-7
  • Electronic_ISBN
    978-1-4244-4134-1
  • Type

    conf

  • DOI
    10.1109/BMEI.2009.5304871
  • Filename