DocumentCode
2184491
Title
A transfer device for 10 V Josephson array measurements
Author
Reymann, D. ; Avrons, D.
Author_Institution
Bur. Int. des Poids et Mesures, Sevres, France
fYear
1994
fDate
June 27 1994-July 1 1994
Firstpage
49
Lastpage
50
Abstract
To improve measurements of 10 V Zener based standards, we have designed a new transfer voltage source for use with 10 V Josephson arrays. The transfer voltage is compared, in turn, with that of the Zener standard and of the array. The use of two separate circuits should help to solve the problem of common ground connections and prevent noise from the Zener standard entering the array circuit. This transfer device can also be used for 10 V Josephson array comparisons.<>
Keywords
Zener effect; measurement standards; superconducting junction devices; voltage measurement; 10 V; Josephson array measurements; Zener standard; array circuit; common ground connections; noise; transfer device; transfer voltage source; Circuit noise; Current measurement; Detectors; Integrated circuit measurements; Measurement standards; Noise measurement; Resistors; Switches; Time measurement; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 1994. Digest., 1994 Conference on
Conference_Location
Boulder, CO, USA
Print_ISBN
0-7803-1984-2
Type
conf
DOI
10.1109/CPEM.1994.333296