Title :
Analysis of the dual heater isothermal AC/DC converter
fDate :
June 27 1994-July 1 1994
Abstract :
The dual heater isothermal AC/DC converter recently introduced by Katzmann (1989) was developed and tested experimentally. This analysis shows that the device used in a negative feedback loop, driven by a sinusoidal voltage, does yield the RMS value and shows the major sources and approximate magnitudes of error.<>
Keywords :
electric heating; electric variables measurement; feedback; measurement errors; measurement standards; thermocouples; Katzmann; Laplace transform; RMS value; dual heater isothermal AC/DC converter; measurement standards; negative feedback loop; sinusoidal voltage; DC-DC power converters; Electric resistance; Equations; Frequency; Isothermal processes; Negative feedback loops; Output feedback; Temperature; Thermal resistance; Voltage;