            Abstract : 
Prior construction of single junction vacuum thermal convertors (SJTCs) did not consistently produce units with AC-DC differences of a few ppm or less. Four improvement modes have been introduced. (1) Elimination of flaming during the making of the mounting bead. (2) Changing of the feed-through posts from Dumet to platinum-iridium. (3) Decrease of the Thomson coefficient by changing from nickel-chromium to Evanohm heater wire. (4) Lowering of the temperature gradient along the heater wire in the vicinity of the thermocouple. SJTCs can now be constructed with AC-DC differences of less than 0.3 ppm at a high yield.<>
         
        
            Keywords : 
Peltier effect; chromium alloys; iridium alloys; measurement standards; nickel alloys; platinum alloys; thermocouples; units (measurement); AC-DC differences; Dumet; Evanohm; Ni-Cr; Pt-Ir; Thomson coefficient; feed-through posts; flaming; heater wire; mounting bead; nickel-chromium; platinum-iridium; temperature gradient; vacuum thermocouples; Australia; Converters; Laboratories; Manufacturing processes; Physics; Production; Temperature; Testing; Welding; Wire;
         
        
        
        