Improved interstitial and intravascular MR imaging coils
Author
Konyer, N.B. ; Viswanathan, R. ; Raghavan, R. ; Mills, G. ; Brady, M. ; Bronskill, M.J.
Author_Institution
Sunnybrook & Women´´s Coll. Health Sci. Centre, Toronto Univ., Ont., Canada
fYear
2002
fDate
2002
Firstpage
963
Lastpage
965
Abstract
An improved intravascular (IV) and interstitial (IS) prototype magnetic resonance (MR) imaging coil design offers dramatic gains in signal-to-noise ratio (SNR) and imaging volume. A multi-turn rectangular solenoid was wound inside catheters with 1.7 mm (IV) and 3.0 mm (IS) diameters. A factor of 200 in peak SNR gain over the standard birdcage head coil was observed, with the region of improved SNR persisting for a radial distance of 10 and 15 mm respectively. Very high resolution ex-vivo intravascular images and very high sensitivity MR spectroscopy (MRS) are demonstrated. This coil design offers interesting performance options in SNR and imaging volume compared to existing FDA-approved catheter coil designs.
Keywords
NMR spectroscopy; biomedical MRI; blood vessels; catheters; solenoids; 1.7 mm; 10 mm; 15 mm; 3.0 mm; catheters; coil design; imaging volume; interstitial MR imaging coils; intravascular MR imaging coils; magnetic resonance imaging coil design; multi-turn rectangular solenoid; peak SNR gain; radial distance; signal-to-noise ratio; very high resolution ex-vivo intravascular images; very high sensitivity MR spectroscopy; Catheters; Coils; Head; Magnetic resonance; Magnetic resonance imaging; Prototypes; Signal design; Signal to noise ratio; Solenoids; Wounds;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Imaging, 2002. Proceedings. 2002 IEEE International Symposium on
Print_ISBN
0-7803-7584-X
Type
conf
DOI
10.1109/ISBI.2002.1029422
Filename
1029422
Link To Document