Abstract
A new method to calibrate thermoelectric power sensors is established and investigated, where the effective efficiency is directly determined in a microcalorimeter. A commercially available power-thermovoltage converter (power sensor) without any electronics is used. The results show good agreement with results obtained by comparing the sensor with a calibrated thermistor mount, and smaller uncertainties.<>
Keywords
calibration; calorimetry; microwave detectors; microwave measurement; power measurement; thermocouples; calibration; effective efficiency; measurement uncertainties; microcalorimetric measurement; microwave power sensors; power sensor; power-thermovoltage converter; thermocouples; thermoelectric power sensors; Bolometers; Calibration; Electronic components; Microwave measurements; Microwave sensors; Power measurement; Resistors; Thermal sensors; Thermoelectricity; Uncertainty;