  • Abstract
    In general, the process of the SPRT module calibration takes long time and requires all employees participation in the calibration process, and it also requires an expert for calculation of results and measurement uncertainty. All of these waste a vast amount of working time. Therefore, this paper provides the automatic SPRT module calibration process which can reduce the steps of calibration, resulting in process time reduction and the automatic process doesn´t require the expert to calculate results and measurement uncertainty.
  • Keywords
    calibration; measurement uncertainty; modules; resistance thermometers; automated low thermal matrix scanner; automatic SPRT module calibration process; measurement uncertainty; standard platinum resistance thermometer; time reduction process; Humans; Thermal resistance;