DocumentCode :
2191415
Title :
Large area gamma-ray imaging detector based on high resolution CdTe diode
Author :
Mitani, Takefumi ; Nakamura, Hidehito ; Uno, Shingo ; Takahashi, Tadayuki ; Nakazawa, Kazuhiro ; Watanabe, Shin ; Tajima, Hiroyasu ; Nomachi, Masaharu ; Fukazawa, Yasushi ; Kubo, Shin ; Kuroda, Yoshikatsu ; Onishi, Mitsunobu ; Ohno, Ryoichi
Author_Institution :
Inst. of Space & Astron. Sci., Kanagawa, Japan
Volume :
1
fYear :
2002
fDate :
10-16 Nov. 2002
Firstpage :
274
Abstract :
We are developing a large array detector composed of 1024 individual CdTe diodes. The each detector has the dimensions of 1.2 mm × 5.0 mm and a thickness of 1.2 mm. An edge-on geometry is used for the injection of gamma-rays. With this geometry, the distance between the two electrodes can be kept small, and we can therefore apply the high electric field which is necessary to achieve the high energy resolution (by reducing the low energy tail) and also to sustain the long-term stability of the CdTe diode. Signals from each detector element are read into newly developed low noise ASICs. We use 32 chips for the readout of 1024 elements. In this paper, we will report basic characteristics of individual detectors and overall performance of the gamma-camera. Design or the readout electronics system is also described.
Keywords :
II-VI semiconductors; cadmium compounds; gamma-ray detection; nuclear electronics; readout electronics; semiconductor counters; CdTe; edge-on geometry; gamma-camera; high electric field; high resolution CdTe diode; large area gamma-ray imaging detector; long-term stability; readout electronics system; Diodes; Electrodes; Energy resolution; Gamma ray detection; Gamma ray detectors; Geometry; Image edge detection; Image resolution; Nuclear imaging; Sensor arrays;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2002 IEEE
Print_ISBN :
0-7803-7636-6
Type :
conf
DOI :
10.1109/NSSMIC.2002.1239315
Filename :
1239315