Title :
Analysis of Vector-ECG and Magneto-ECG Based on MCG Data
Author :
Gu, Qingwen ; Jiang, Shiqin ; Wang, Weiyuan ; Quan, Weiwei
Author_Institution :
Tongji Univ., Shanghai
Abstract :
A method for obtaining the Magneto-ECG (M-ECG) from MCG Data was proposed by our previous research. In present paper, the related error of this method is analyzed by using a single current dipole model in simulation. The aim is to compare the calculated M-ECG with the recorded ECG and provide a simple method to estimate the myocardial current. Our result indicates that the calculated vector electrocardiogram (V-ECG) is available when an equivalent dipole is under the central area of detecting surface. Furthermore, a comparison of the calculated lead-II M-ECG and the synchronously recorded lead-II ECG is shown from 20 subjects.
Keywords :
electrocardiography; magnetocardiography; MCG; electrocardiogram; magneto-ECG; myocardial current; single current dipole model; vector electrocardiogram; vector-ECG; Cardiac disease; Data engineering; Electrocardiography; Heart; Humans; Information analysis; Magnetic analysis; Magnetic field measurement; Magnetic fields; Myocardium; Biomagnetics; ECG; MCG; vector ECG;
Conference_Titel :
Noninvasive Functional Source Imaging of the Brain and Heart and the International Conference on Functional Biomedical Imaging, 2007. NFSI-ICFBI 2007. Joint Meeting of the 6th International Symposium on
Conference_Location :
Hangzhou
Print_ISBN :
978-1-4244-0949-5
Electronic_ISBN :
978-1-4244-0949-5
DOI :
10.1109/NFSI-ICFBI.2007.4387700