• DocumentCode
    2194310
  • Title

    Performance measurements of a pixelated NaI(Tl) PET scanner

  • Author

    Perkins, A.E. ; Muehllehner, G. ; Surti, S. ; Karp, J.S.

  • Author_Institution
    Philips Med. Syst., Philadelphia, PA, USA
  • Volume
    2
  • fYear
    2002
  • fDate
    10-16 Nov. 2002
  • Firstpage
    885
  • Abstract
    A prototype PET scanner for whole body imaging with 4 mm × 4 mm × 30 mm NaI(Tl) crystals and Anger-logic readout has been built and tested. The scanner is composed of 36,540 NaI(Tl) pixels which are coupled to an optically continuous lightguide and a hexagonal closed packed array of 39 mm photomultiplier tubes (PMTs). The scanner is designed with a crystal-to-crystal diameter of 89 cm and an axial field of view (AFOV) of 25 cm. The main goals of this study are (1) to overcome the count-rate limitation of the continuous NaI(Tl) scanner (CPET), (2) to improve the spatial resolution and image contrast by using small pixels, and (3) to eliminate the relatively large data gaps between the detectors in the continuous NaI(Tl) scanner.
  • Keywords
    photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; 30 mm; 4 mm; 89 cm; Anger-logic readout; NaI:Tl; performance measurements; photomultiplier tubes; pixelated NaI(Tl) PET scanner; small pixels; whole body imaging; Crystals; Measurement; Optical arrays; Optical coupling; Photomultipliers; Positron emission tomography; Prototypes; Testing; Whole body imaging; Whole-body PET;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2002 IEEE
  • Print_ISBN
    0-7803-7636-6
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2002.1239466
  • Filename
    1239466