Abstract :
This paper estimate the magnitude and phase response of a sampling oscilloscope (Agilent86100C with 50 GHz bandwidth) using the nose-to-nose method (NTN). The measurements are corrected for the non-ideal properties of the oscilloscope and calibration apparatus. In this paper, we present the first published repeatability study of magnitude and phase responses measured by NTN method and present the magnitude and phase expanded uncertainty of the NTN method
Keywords :
frequency estimation; frequency response; millimetre wave devices; oscilloscopes; sampling methods; 50 GHz; NTN method; frequency response estimation; nose-to-nose method; sampling oscilloscope; Calibration; Circuits; Frequency estimation; Frequency measurement; Frequency response; Oscilloscopes; Phase measurement; Pulse measurements; Sampling methods; Uncertainty;