• DocumentCode
    2201940
  • Title

    Measurement Errors Caused by the Transient Limiter

  • Author

    Williams, Tim

  • Author_Institution
    Elmac Services, Wareham, United Kingdom
  • fYear
    2007
  • fDate
    24-28 Sept. 2007
  • Firstpage
    493
  • Lastpage
    496
  • Abstract
    The standard conducted emissions test may use a transient limiter between the LISN and the measuring instrumentation. This component is non-linear and can introduce significant errors in the measurement, sufficient to completely change the outcome of the test but which are transparent to the test engineer. The different mechanisms by which this may happen are described and modelled, and some mitigation techniques are recommended.
  • Keywords
    Attenuation; Clamps; Diodes; Impedance; Instruments; Measurement errors; Protection; Radio frequency; Testing; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Compatibility, 2007. EMC Zurich 2007. 18th International Zurich Symposium on
  • Conference_Location
    Munich, Germany
  • Print_ISBN
    978-3-9523286-1-3
  • Electronic_ISBN
    978-3-9523286-0-6
  • Type

    conf

  • DOI
    10.1109/EMCZUR.2007.4388303
  • Filename
    4388303