Title :
Wireless Vestibular Evoked Myogenic Potentials System
Author :
Torfs, T. ; Yazicioglu, R.F. ; Merken, P. ; Gyselinckx, B. ; Puers, R. ; Vanspauwen, R. ; Wuyts, F.L. ; Van Hoof, C.
Author_Institution :
IMEC, Leuven
Abstract :
Existing commercial vestibular evoked myogenic potentials (VEMP) testing systems are cabled systems, which impede certain experiments, particularly those that involve motion and rotation of the patient. This paper presents an autonomous wireless system to record VEMPs. The system uses IMEC´s 60 muW 60 nV/VHz biopotential readout front-end to extract the electromyogram (EMG). It uses IMEC´s low power processing and wireless platform to perform all signal processing locally, significantly reducing the amount of wireless data transmission. The system also implements a low power audio synthesizer and an amplifier to create the necessary audio or vibration stimuli that are applied to the patient. The system can perform 1000 measurements on 2 AA batteries.
Keywords :
amplifiers; biomedical telemetry; electromyography; EMG; amplifier; cabled systems; electromyogram; power audio synthesizer; vestibular evoked myogenic potentials testing; wireless vestibular evoked myogenic potentials system; Aerospace testing; Electrodes; Electromyography; Humans; Irrigation; Measurement standards; Muscles; Neural pathways; System testing; Vibration measurement;
Conference_Titel :
Sensors, 2007 IEEE
Conference_Location :
Atlanta, GA
Print_ISBN :
978-1-4244-1261-7
Electronic_ISBN :
1930-0395
DOI :
10.1109/ICSENS.2007.4388600