Title :
Multiwavelength transmission spectrophotometry in the pulsatile measurement of hemoglobin derivatives in whole blood
Author :
Morillo, E.C. ; Mendelson, Y.
Author_Institution :
Dept. of Electr. Eng., Worcester Polytech. Inst., MA, USA
fYear :
1997
fDate :
21-22 May 1997
Firstpage :
5
Lastpage :
6
Abstract :
This paper describes a multiwavelength transmission spectrophotometry set up to study the feasibility of measuring hemoglobin derivatives in whole blood, utilizing transmission pulse oximetry. Comparison of measured relative absorbance spectra with standard extinction coefficients versus wavelength data shows that the system closely reproduces the standard spectra of blood
Keywords :
blood flow measurement; proteins; pulsatile flow; spectrochemical analysis; spectrophotometry; hemoglobin derivatives; in-vitro blood circulation system; multiwavelength transmission spectrophotometry; photoplethysmographic pulse; pulsatile measurement; relative absorbance spectra; standard extinction coefficients; transmission pulse oximetry; whole blood; Biomedical measurements; Blood; Fingers; High speed optical techniques; Holographic optical components; Holography; Optical filters; Optical pulses; Optical saturation; Pulse measurements;