Our goal was to record electrocardiogram (ECG) and respiration data from an awake, freely moving rat. The technique presented here requires simple surgery in which two electrodes and a ground wire are placed subcutaneously across the chest, which is externalized via a connector on the top of the subjects head. From these leads it is possible to collect ECG in a standard two lead configuration, and respiration with impedance pneumography. Utilizing electrode placement and signal filtering we were able to obtain ECG from anaesthetized, and awake rats without the effects of motion artifact. Respiration was collected with little effect of motion in all but one subject which displayed excessive movement
Keywords
biological techniques; biology computing; electric impedance measurement; electrocardiography; flow measurement; medical signal processing; pneumodynamics; surgery; ECG data; awake freely moving rat; awake unrestrained rat; chest; connector; electrocardiogram; electrode placement; electrodes; ground wire; head; impedance pneumography; instrumentation; motion artifact; respiration data; signal filtering; simple surgery; standard two lead configuration; Connectors; Electrocardiography; Electrodes; Filtering; Impedance; Instruments; Magnetic heads; Rats; Surgery; Wire;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 1997., Proceedings of the IEEE 1997 23rd Northeast
Conference_Location
Durham, NH
Print_ISBN
0-7803-3848-0
Type
conf
DOI
10.1109/NEBC.1997.594942
Filename
594942
Link To Document