DocumentCode :
2212203
Title :
Contactless calibration of loop antennas in comparison to traditional methods
Author :
Kullmer, A. ; Richter, N. ; Lahme, M. ; Enders, A. ; Junge, A.
Author_Institution :
Institute for Electromagnetic compatibility, TU Braunschweig, Germany
fYear :
2015
fDate :
16-22 Aug. 2015
Firstpage :
876
Lastpage :
880
Abstract :
Magnetic loop antennas are commonly used for investigations on magnetic fields in a frequency range up to 30 MHz. As for all antennas the antenna factor is the result of a calibration procedure. Commonly used calibration routines like the standard field, standard antenna and three antenna method will be compared to a newer approach called impedance method, first introduced by Ishii [2]. Finally a new approach for contactless measurement of the antenna factor, the so-called in-situ impedance method, will be presented as well as a comparison of the resulting antenna factors of all calibration routines.
Keywords :
Antenna measurements; Antennas; Calibration; Impedance; Impedance measurement; Magnetic field measurement; Standards; calibration; electromagnetic compatibility; electromagnetic fields; loop antenna; test equipment;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility (EMC), 2015 IEEE International Symposium on
Conference_Location :
Dresden, Germany
Print_ISBN :
978-1-4799-6615-8
Type :
conf
DOI :
10.1109/ISEMC.2015.7256280
Filename :
7256280
Link To Document :
بازگشت