DocumentCode :
2219875
Title :
Precision of glucose measurement in Intralipid suspensions with optical coherence tomography
Author :
Feng, Shu ; Yuan, Kehong ; Ye, Datian
Author_Institution :
Grad. Sch. at Shenzhen&Dept. of Biomed. Eng., Tsinghua Univ., Tsinghua
fYear :
2008
fDate :
30-31 May 2008
Firstpage :
510
Lastpage :
513
Abstract :
Optical coherence tomography (OCT) presents its capability of noninvasive glucose measurement in tissue-simulating phantoms and biological tissues. However, speckle noise substantially deteriorates the accuracy of the measurements with this technique. The speckle noise can be suppressed by averaging of in-depth scans. We studied the suppression of speckle noise for accurate measurement of backscattering signal and the maximum precision of glucose measurement in Intralipid suspensions with the OCT technique. Our results show that the precision of glucose measurement can achieve plusmn4.4 mM for 10% Intralipid and plusmn2.2 mM for 3% Intralipid after the averaging of 50,000 scans.
Keywords :
biochemistry; biological tissues; chemical variables measurement; optical tomography; phantoms; speckle; backscattering signal measurement; biological tissues; glucose measurement; intralipid suspensions; optical coherence tomography; speckle noise suppression; tissue-simulating phantoms; Adaptive optics; Biological tissues; Biomedical optical imaging; Imaging phantoms; Noise measurement; Optical noise; Speckle; Sugar; Suspensions; Tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Information Technology and Applications in Biomedicine, 2008. ITAB 2008. International Conference on
Conference_Location :
Shenzhen
Print_ISBN :
978-1-4244-2254-8
Electronic_ISBN :
978-1-4244-2255-5
Type :
conf
DOI :
10.1109/ITAB.2008.4570660
Filename :
4570660