  • Abstract
    This paper investigates design, modelling, simulation, and control issues related to Doppler ultrasound involving an embedded system. The aim is to view the Doppler spectrogram commonly used in clinical ultrasound and compute the heart rate and blood velocity in the scanned part. Also, it is possible to detect the presence of turbulent or laminar flow. We introduce the embedded ultrasound imaging device based on the Beagleboard and allow the computations to be distributed within the integrated DM3730 processor beyween its ARM and DSP processors. The resulting system is low in cost, small in size and provide an interface that is comparable to PC-based systems indicating potential for clinical utility.
  • Keywords
    Doppler measurement; biomedical ultrasonics; blood flow measurement; cardiology; laminar flow; turbulence; velocity measurement; ARM processor; DM3730 processor; DSP processor; Doppler spectrogram; blood velocity; embedded real time ultrasound system; heart rate and; laminar flow; turbulent flow; Digital signal processing; Doppler effect; Educational institutions; Graphical user interfaces; Real time systems; Spectrogram; Ultrasonic imaging; Doppler ultrasound; digital signal processing; embedded systems;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Radio Science Conference (NRSC), 2012 29th National
  • Conference_Location
    Cairo
  • Print_ISBN
    978-1-4673-1884-6
  • Type

    conf

  • DOI
    10.1109/NRSC.2012.6208585
  • Filename
    6208585