• DocumentCode
    2232181
  • Title

    B-type imaging with coded signals

  • Author

    Gindre, M. ; Urbach, W.

  • fYear
    1993
  • fDate
    31 Oct-3 Nov 1993
  • Firstpage
    1171
  • Abstract
    We have used a pseudorandom (PN) coded signal to investigate the improvement of the signal to noise ratio (SNR) in B-type echography. With N≈103 bits of code the experimental value of SNR is about 55 dB for PN sequence, close to the theoretical value of 20 log N ≈60 dB. In order to assess the respective advantages of the classical echo images and those obtained with coded signal, we present some B-type images obtained from objects of elementary geometry with application to non destructive testing and medical imaging. We show that the SNR improvement obtained with PN sequence allows the observation of details which would remain invisible with classical echographic techniques
  • Keywords
    acoustic imaging; biomedical ultrasonics; echo; patient diagnosis; ultrasonic materials testing; ultrasonic reflection; 55 dB; B-type echography; B-type images; classical echo images; coded signal; medical imaging; nondestructive testing; pseudorandom coded signal; signal to noise ratio; Bandwidth; Biomedical imaging; Energy resolution; Frequency; Instruments; Medical diagnostic imaging; Medical tests; Phase modulation; Signal to noise ratio; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1993. Proceedings., IEEE 1993
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    0-7803-2012-3
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1993.339624
  • Filename
    339624