• DocumentCode
    2238990
  • Title

    Closed loop stability analysis of an assistance system for catheterization

  • Author

    Kern, Thorsten A. ; Klages, Stephanie ; Meiß, Thorsten ; Werthschützky, Roland

  • Author_Institution
    Inst. for Electromech. Design, Darmstadt Univ. of Technol.
  • fYear
    2007
  • fDate
    22-24 March 2007
  • Firstpage
    597
  • Lastpage
    598
  • Abstract
    Catheterizations are standard procedures of today´s vascular treatment for diagnosis and therapy. To improve navigation even into smaller branches a novel device with haptic feedback assisting during catheterization is currently being developed. One aspect of the design is the identification of possible sources for instability, which is focused on in this paper. This requires the analysis and modeling of the interaction between the device, the patient and the medical staff. The resulting open loop response plotted in Bode-diagram demonstrates areas of possible instabilities. As a result design constrains to guarantee stability in any possible situation is derived
  • Keywords
    Bode diagrams; closed loop systems; force feedback; haptic interfaces; patient diagnosis; Bode-diagram; catheterization assistance system; closed loop stability analysis; haptic feedback; open loop response; patient diagnosis; patient therapy; vascular treatment; Actuators; Catheterization; Frequency; Haptic interfaces; Medical diagnostic imaging; Medical treatment; Navigation; Stability analysis; Veins; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    EuroHaptics Conference, 2007 and Symposium on Haptic Interfaces for Virtual Environment and Teleoperator Systems. World Haptics 2007. Second Joint