Title :
A calibration procedure for computer controlled multichannel pulse-height analyzer
Author :
Cacciari, M. ; Contin, A.
Author_Institution :
Dipartimento di Elettrotecnica Elettronica ed Inf., Trieste Univ., Italy
Abstract :
The availability of a Computer Controlled Multichannel Analyzer (MCA) Measurement System, for partial discharge (PD) measurements, opens interesting possibilities to develop software tools devoted to assist the calibration procedures. In this paper, a calibration routine designed to establish a correlation between the quantization level of the A/D converter and the apparent charge quantity in a complete test circuit, is presented and discussed. The proposed procedure is based on statistical processing of the calibration signals produced accordingly to IEC Std (1) or ASTM Std (2). In order to evaluate the correctness of the calibration results, some calibration factors are proposed
Keywords :
calibration; computerised instrumentation; partial discharges; pulse height analysers; quantisation (signal); A/D converter; calibration; charge quantity; computer controlled multichannel pulse-height analyzer; partial discharge measurements; quantization; software tools; statistical signal processing; test circuit;
Conference_Titel :
Partial Discharge, 1993., International Conference on
Conference_Location :
Canterbury
Print_ISBN :
0-85296-579-6