• DocumentCode
    2247
  • Title

    Design and Evaluation of a Quasi-Passive Knee Exoskeleton for Investigation of Motor Adaptation in Lower Extremity Joints

  • Author

    Shamaei, Kamran ; Cenciarini, Massimo ; Adams, Albert A. ; Gregorczyk, Karen N. ; Schiffman, Jeffrey M. ; Dollar, Aaron M.

  • Author_Institution
    Dept. of Mech. Eng. & Mater. Sci., Yale Univ., New Haven, CT, USA
  • Volume
    61
  • Issue
    6
  • fYear
    2014
  • fDate
    Jun-14
  • Firstpage
    1809
  • Lastpage
    1821
  • Abstract
    In this study, we describe the mechanical design and control scheme of a quasi-passive knee exoskeleton intended to investigate the biomechanical behavior of the knee joint during interaction with externally applied impedances. As the human knee behaves much like a linear spring during the stance phase of normal walking gait, the exoskeleton implements a spring across the knee in the weight acceptance (WA) phase of the gait while allowing free motion throughout the rest of the gait cycle, accomplished via an electromechanical clutch. The stiffness of the device is able to be varied by swapping springs, and the timing of engagement/disengagement changed to accommodate different loading profiles. After describing the design and control, we validate the mechanical performance and reliability of the exoskeleton through cyclic testing on a mechanical knee simulator. We then describe a preliminary experiment on three healthy adults to evaluate the functionality of the device on both left and right legs. The kinetic and kinematic analyses of these subjects show that the exoskeleton assistance can partially/fully replace the function of the knee joint and obtain nearly invariant moment and angle profiles for the hip and ankle joints, and the overall knee joint and exoskeleton complex under the applied moments of the exoskeleton versus the control condition, implying that the subjects undergo a considerable amount of motor adaptation in their lower extremities to the exoskeletal impedances, and encouraging more in-depth future experiments with the device.
  • Keywords
    biomedical equipment; bone; elastic constants; gait analysis; kinematics; ankle joints; biomechanical behavior; device stiffness; electromechanical clutch; exoskeleton implements; extremity joints; gait cycle; hip joints; human knee; kinematic analyses; kinetic analyses; knee joint; left legs; mechanical design; mechanical knee simulator; mechanical performance; motor adaptation; normal walking gait; quasi-passive knee exoskeleton; right legs; stance phase; swapping springs; weight acceptance phase; Exoskeletons; Friction; Joints; Knee; Shafts; Springs; Knee biomechanics; lower extremity exoskeleton; quasi-passive mechanism; variable-stiffness;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2014.2307698
  • Filename
    6747325