• DocumentCode
    2254168
  • Title

    Algorithm for calibration of the electromagnetic tracking system

  • Author

    Bien, Tomasz ; Rose, Georg

  • Author_Institution
    Dept. of Electr. Eng., Otto-von-Guericke Univ., Magdeburg, Germany
  • fYear
    2012
  • fDate
    5-7 Jan. 2012
  • Firstpage
    85
  • Lastpage
    88
  • Abstract
    Electromagnetic tracking systems (EMTS) are increasingly used in computer assisted surgery systems, motion capture units and military systems. The main application in the medical domain represents the support of minimally invasive surgeries. However, those electromagnetic tracking systems suffer from low accuracy and systematic errors caused by conductive and ferromagnetic media. In this paper we present a method for the calibration of electromagnetic tracking systems, which could be adopted in the development of novel electromagnetic tracking devices. The proposed method was evaluated on an experimental setup of a tracking system.
  • Keywords
    calibration; electromagnetic devices; measurement errors; surgery; tracking; calibration algorithm; computer assisted surgery systems; electromagnetic tracking devices; electromagnetic tracking system; ferromagnetic media; medical application; military systems; minimally invasive surgeries; motion capture units; systematic errors; Accuracy; Biomechanics; Computers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical and Health Informatics (BHI), 2012 IEEE-EMBS International Conference on
  • Conference_Location
    Hong Kong
  • Print_ISBN
    978-1-4577-2176-2
  • Electronic_ISBN
    978-1-4577-2175-5
  • Type

    conf

  • DOI
    10.1109/BHI.2012.6211512
  • Filename
    6211512