    Wrist cuff method determines mean arterial pressure in dual-cuff blood pressure system

  • Author

    Jilek, Jiri ; Stork, Milan

  • Author_Institution
    CarditechCulver City, Culver City, CA, USA
  • fYear
    2014
  • fDate
    9-10 Sept. 2014
  • Firstpage
    145
  • Lastpage
    148
  • Abstract
    Mean arterial pressure (MAP) is the pressure in the arteries averaged over time. Oscillometric blood pressure (BP) monitors determine the value of MAP by processing arterial pulses in the cuff. Cuff pressure at the point of largest pulse amplitude is equal to MAP. The oscillometric determination of MAP is subject to errors resulting from wave reflections and stroke volume variability. We developed an experimental system that uses two cuffs in order to improve accuracy of noninvasive BP measurement. The system consists of a compact module with pneumatic and electronic circuits, two detachable cuffs (arm and wrist), and a notebook computer connected to the module via a USB cable. This study concentrated on improvements in the determination of MAP using the wrist cuff pulses. The smallest wrist cuff amplitude in the middle segment of BP test was determined as the point of MAP. The corresponding arm cuff pressure was the value of MAP. Waveforms and cuff pressures from 40 BP tests were used to determine oscillometric, wrist cuff, and calculated MAP values and the values were compared. The results indicated that it is possible to determine MAP from wrist cuff pulse amplitudes and that the wrist cuff method may be more accurate than the oscillometric method.
  • Keywords
    biomedical equipment; blood pressure measurement; blood vessels; patient monitoring; waveform analysis; MAP point; MAP value calculation; USB cable; arm cuff pressure; arterial pulse processing; compact module; cuff pressures; detachable arm cuff; detachable wrist cuff; dual-cuff blood pressure system; electronic circuit; mean arterial pressure determination; middle BP test segment; noninvasive BP measurement accuracy; notebook computer; oscillometric BP monitor; oscillometric MAP determination error; oscillometric MAP value determination; oscillometric blood pressure monitor; pneumatic circuit; pulse amplitude point; small wrist cuff amplitude determination; stroke volume variability; wave reflection; waveform; wrist cuff MAP value determination; wrist cuff method; wrist cuff pulse amplitude; Biomedical monitoring; Blood pressure; Pressure measurement; Reflection; Universal Serial Bus; Wrist; Mean arterial pressure; blood pressure; dual cuff system; oscillometric method; wave reflections; wrist cuff pulse;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Applied Electronics (AE), 2014 International Conference on
  • Conference_Location
    Pilsen
  • ISSN
    1803-7232
  • Print_ISBN
    978-8-0261-0276-2
  • Type

    conf

  • DOI
    10.1109/AE.2014.7011687
  • Filename
    7011687