• DocumentCode
    2275707
  • Title

    Extending ac-dc current transfer measurement to 100 A, 100 kHz

  • Author

    Rydler, Karl-Erik ; Tarasso, Valter

  • Author_Institution
    SP Tech. Res., Inst. of Sweden, Boras
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    28
  • Lastpage
    29
  • Abstract
    This paper describes the design and verification of current shunts for AC-DC transfer measurements in the range 20 A to 100 A. The measured AC-DC current transfer difference of the 100 A shunt is -36 muA/A at 100 kHz with an uncertainty of 200 muA/A.
  • Keywords
    electric current measurement; measurement uncertainty; transfer standards; AC-DC current transfer measurement; current 20 A to 100 A; current shunt; frequency 100 kHz; uncertainty measurement; Analog-digital conversion; Circuits; Coaxial components; Conductors; Current measurement; Frequency; Measurement standards; Mutual coupling; Resistors; Ribs;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
  • Conference_Location
    Broomfield, CO
  • Print_ISBN
    978-1-4244-2399-6
  • Electronic_ISBN
    978-1-4244-2400-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2008.4574636
  • Filename
    4574636