Title :
Performances and progress of a compact pulsed CPT Cs clock
Author :
Guerandel, S. ; Boudot, R. ; Castagna, N. ; Clercq, E. De ; Dimarcq, N. ; Clairon, A.
Author_Institution :
Obs. de Paris, SYRTE, Paris
Abstract :
We report on the development of a compact Cs cell CPT frequency standard operating in the Dicke regime. Combining a double Lambda scheme and a Ramsey interrogation, high contrast fringes whose linewidth depends on the time separating two pulses, are obtained. Compared to the continuous regime, the pulsed interrogation reduces the light shift by a factor 300 and permits to reach frequency stability 10 times better. A frequency stability of 9.10-13 at 1 s has already been measured.
Keywords :
atomic clocks; frequency standards; laser frequency stability; radiation pressure; Dicke regime; Ramsey interrogation; compact pulsed CPT caesium clock; contrast fringes; double Lambda scheme; frequency stability; frequency standard; pulsed interrogation; Atomic beams; Clocks; Frequency; Gas lasers; Laser beams; Laser excitation; Laser noise; Masers; Stability; Stationary state;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
DOI :
10.1109/CPEM.2008.4574654