• DocumentCode
    2277657
  • Title

    Study on the calibration of partial discharge calibrator

  • Author

    Na, Liu ; Zhi, Li ; Zheng, Qian

  • Author_Institution
    Electr. Magn. Div., Nat. Inst. of Metrol., Beijing
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    232
  • Lastpage
    233
  • Abstract
    The calibration of apparent charge, which can be traced to AC voltage and capacitance, for partial discharge(PD) calibrator is discussed. The methods of fast sampling measurement of peak-to-peak amplitude for step voltage by digital multimeter and the arithmetic in the automatic measurement of apparent charge by digital oscilloscope are focused on. Then, other factors, such as impulse waveform and the choice of capacitor are also discussed.
  • Keywords
    calibration; digital multimeters; oscilloscopes; partial discharge measurement; digital multimeter; digital oscilloscope; partial discharge calibrator; sampling measurement; Calibration; Capacitance; Charge measurement; Current measurement; Digital arithmetic; Oscilloscopes; Partial discharge measurement; Partial discharges; Sampling methods; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
  • Conference_Location
    Broomfield, CO
  • Print_ISBN
    978-1-4244-2399-6
  • Electronic_ISBN
    978-1-4244-2400-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2008.4574738
  • Filename
    4574738