• DocumentCode
    2278000
  • Title

    A new synchronous method for attenuation measurement system

  • Author

    Gao Qiulai ; Liang Weijun ; Huang Hui ; Li Yong

  • Author_Institution
    Nat. Inst. of Metrol., Beijing
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    262
  • Lastpage
    263
  • Abstract
    Audio frequency substitution with inductive voltage divider (IVD) as reference standard is the most important attenuator calibration system in which a lock-in amplifier is widely used as phase sensitive detector. In order to obtain synchronous signal for lock-in amplifier, a new simple system structures with good stability was developed in NIM (China). Its principle and experiment data are discussed in the paper.
  • Keywords
    attenuation measurement; calibration; intermediate-frequency amplifiers; measurement systems; stability; voltage dividers; attenuation measurement system; attenuator calibration system; audio frequency substitution; inductive voltage divider; lock-in amplifier; phase sensitive detector; stability; synchronous method; synchronous signal; Amplifiers; Attenuation measurement; Frequency conversion; Frequency modulation; Measurement standards; Microwave measurements; Phase detection; Stability; Testing; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
  • Conference_Location
    Broomfield, CO
  • Print_ISBN
    978-1-4244-2399-6
  • Electronic_ISBN
    978-1-4244-2400-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2008.4574753
  • Filename
    4574753