DocumentCode :
2278118
Title :
Novel digital voltage ramp generator for picoammeter calibrations
Author :
Willenberg, G.D. ; Tauscher, H.N.
Author_Institution :
Phys.-Tech. Bundesanstalt, Braunschweig
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
276
Lastpage :
277
Abstract :
For the calibration of picoammeters a novel highly linear voltage ramp generator has been constructed which is based on digital electronics and two digital-to-analog-converters (DACs). In contrast to our previous ramp generator constructed entirely in analog electronics the new design allows for the first time to perform dynamic measurements without any sacrifice of linearity.
Keywords :
ammeters; calibration; digital-analogue conversion; ramp generators; calibration; digital electronics; digital voltage ramp generator; digital-analog converters; dynamic measurements; linear voltage ramp generator; picoammeter; Calibration; Capacitance; Capacitors; Linearity; Microcontrollers; Performance evaluation; Prototypes; Stability; Time measurement; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574760
Filename :
4574760
Link To Document :