Abstract :
An accurate RF attenuation measurement, with small mismatch effect, is achieved without source and load being matched to the line impedance. Insertion loss measurements to the DUT and to the DUT fitted by quarter-wavelength lines are carried out, then the attenuation is obtained simply, unrelated to the source and load reflection coefficients. Experiment results on two 20-dB attenuators at 11 GHz show good agreements with those being impedance matched.
Keywords :
attenuation measurement; impedance matching; loss measurement; DUT; device under test; insertion loss measurement; line impedance matching; load reflection coefficient; mismatch effect; precision RF attenuation measurement; quarter-wavelength lines; Attenuation measurement; Attenuators; Impedance measurement; Insertion loss; Loss measurement; Measurement standards; Metrology; Microwave measurements; Radio frequency; Reflection;