Abstract
This paper describes an DC voltage calibration system which is improved to be more accurate and more easy to operate. It is based on 10 V Josephson array, and uses a fully automated system consisting of switches setup well designed for accurate measurement. An overall standard uncertainty is estimated less than 5 nV(1sigma) at 10 V level.
Keywords
Josephson effect; calibration; measurement standards; measurement uncertainty; superconducting arrays; voltage measurement; DC reference; DC voltage calibration system; Josephson array; automated calibration system; measurement uncertainty; voltage 10 V; Automatic control; Calibration; Control systems; Diodes; Engines; Metrology; Phased arrays; Switches; Uncertainty; Voltage;