• DocumentCode
    2281266
  • Title

    Precision ac-dc transfer measurement system based on a 1000-volt inductive voltage divider

  • Author

    Budovsky, I. ; Hagen, T.

  • Author_Institution
    Nat. Meas. Inst., Lindfield, NSW
  • fYear
    2008
  • fDate
    8-13 June 2008
  • Firstpage
    656
  • Lastpage
    657
  • Abstract
    The paper presents a new AC-DC transfer measurement system for voltages up to 1000 V and frequencies up to 1 kHz based on a precision inductive voltage divider. Preliminary evaluations show the possibility of standard uncertainties of less than 3 muV/V at 1000 V.
  • Keywords
    calibration; measurement systems; measurement uncertainty; transfer standards; voltage dividers; voltage measurement; calibration; high-voltage AC-DC difference measurement; inductive voltage divider; measurement uncertainty; precision AC-DC transfer measurement system; voltage 1000 V; voltage measurement; Australia; Cables; Calibration; Frequency measurement; Power amplifiers; Relays; Switches; System testing; Uncertainty; Voltage;