Title :
The calibration of IEC standard flickermeters using complex modulated signals
Author :
Clarkson, P. ; Wright, P.S.
Author_Institution :
Nat. Phys. Lab., Teddington, UK
Abstract :
Flickermeter calibrations currently rely on the predicted response of a standard flickermeter to basic waveforms. A description is given of an interpretation and implementation of the IEC standard flickermeter, developed at NPL, in order to assess the response of commercial flickermeters to more complex waveforms, which are representative of signals that occur in practice. A description of the incorporation of the resulting model into a calibration system for flickermeters under complex-waveform conditions is given.
Keywords :
calibration; electric variables measurement; signal processing equipment; waveform analysis; IEC standard flickermeter; calibration; complex modulated signals; complex waveforms; Calibration; Current supplies; Frequency modulation; Home appliances; IEC standards; Laboratories; Manufacturing; Phase modulation; System testing; Voltage fluctuations;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
DOI :
10.1109/CPEM.2008.4574957