DocumentCode
2281441
Title
An experiment for verification of auxiliary transmission line method to evaluate a microwave thermal noise source
Author
Nakano, H. ; Murakami, H. ; Inoue, T. ; Kato, Y.
Author_Institution
Nat. Metrol. Inst. of Japan, Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
fYear
2002
fDate
16-21 June 2002
Firstpage
72
Lastpage
73
Abstract
In order to verify the usefulness of an auxiliary transmission line method, which we have already proposed to evaluate the noise temperature of a microwave thermal noise source, a liquid nitrogen cooled coaxial noise source system has been constructed and its preliminary experiment has been conducted. Then, this measurement result and a result calibrated to the noise source with two known standards have been compared.
Keywords
electric noise measurement; high-frequency transmission lines; microwave measurement; thermal noise; transmission line theory; N/sub 2/; auxiliary transmission line method verification; liquid nitrogen cooled coaxial noise source system; microwave thermal noise source; noise source calibration; noise temperature; standards; Circuit noise; Coaxial components; Microwave theory and techniques; Noise generators; Noise measurement; Power transmission lines; Radiometry; Signal to noise ratio; Temperature; Transmission line measurements;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location
Ottawa, Ontario, Canada
Print_ISBN
0-7803-7242-5
Type
conf
DOI
10.1109/CPEM.2002.1034725
Filename
1034725
Link To Document