DocumentCode :
2281476
Title :
Minimum phase response reconstruction of sampling oscilloscopes based on the NTN calibration
Author :
Qinghua, Xu ; Maoliu, Lin ; Zhe, Zhang
Author_Institution :
Dept. of Inf. Eng., Harbin Inst. of Technol., Harbin, China
fYear :
2008
fDate :
8-13 June 2008
Firstpage :
682
Lastpage :
683
Abstract :
This paper describes an algorithm for determining the minimum phase response of sampling oscilloscopes of a linear time-invariant response function from its magnitude. The procedure is based on Kramers-Kronig relations in combination with the harmonic phase response measurement of the NTN calibration. Although the truncation of the Hilbert transform gives rise to large errors in estimated phase, these errors may be approximated by using three basis functions. As an example, we got the minimum-phase response reconstruction of sampling oscilloscopes in the frequency domain; its frequency resolution achieves 1 MHz.
Keywords :
Hilbert transforms; Kramers-Kronig relations; calibration; oscilloscopes; Hilbert transform; Kramers-Kronig relations; NTN calibration; basis functions; harmonic phase response measurement; linear time-invariant response function; minimum phase response; sampling oscilloscopes; Calibration; Frequency measurement; Least squares approximation; Merging; Oscilloscopes; Phase estimation; Phase measurement; Sampling methods; Signal analysis; Signal resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
Type :
conf
DOI :
10.1109/CPEM.2008.4574963