Title :
Attenuation measurement system in frequency range of 10-100 MHz
Author :
Widarta, A. ; Kawakami, Tomoya
Author_Institution :
AIST, Nat. Metrol. Inst. of Japan (NMIJ), Tsukuba, Japan
Abstract :
An attenuation measurement system, which employs a traceable inductive voltage divider (IVD) at 1 kHz as a reference standard and the dual channel intermediate frequency (IF) substitution method, is developed for an attenuation standard in the frequency range of 10-100 MHz. The basic properties of the system were experimentally investigated at a frequency of 60 MHz.
Keywords :
VHF radio propagation; attenuation measurement; measurement standards; measurement systems; voltage dividers; 1 kHz; 10 to 100 MHz; 60 MHz; attenuation measurement system; attenuation standard; dual channel intermediate frequency substitution method; frequency range; traceable inductive voltage divider reference standard; Assembly; Attenuation measurement; Frequency measurement; Light emitting diodes; Measurement standards; Phase detection; Phase shifters; Radio frequency; Standards development; Voltage;
Conference_Titel :
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location :
Ottawa, Ontario, Canada
Print_ISBN :
0-7803-7242-5
DOI :
10.1109/CPEM.2002.1034729